SPEEDY positions:
Starting with the upstream hemisphere and the top detector and count clockwise:
#1
#4 X #2
#3
with X being the beam into the page.
Correspondingly for the downstream hemisphere:
#5
#8 X #6
#7
SASSYER positions:
As numbered at the array.
Detector E1
Serial Number: 06
Nominal Bias: +3000 V
Design: Type I
Bias shutdown: ORTEC
Status:
|
Date |
Location |
Position |
Problems
|
REaction |
Comments |
Initials |
|
|
SPEEDY |
#3 |
- |
- |
- |
A.H. |
|
|
SASSYER |
#12 |
- |
- |
- |
A.H. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Detector E2
Serial Number: 07
Nominal Bias: +3000V
Design: Type I
Bias shutdown: Ortec
Status:
|
Date |
Location |
Position |
Problems
|
REACTION |
Comments |
Initials |
|
|
SPEEDY |
#2 |
- |
- |
- |
A.H. |
|
|
Hospital |
- |
- |
- |
Tom changes bias shutdown from Eurosys to Ortec. Finished in the evening. |
A.H. |
|
|
SASSYER |
#4 |
- |
- |
- |
A.H. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Detector E3
Serial Number: 10
Nominal Bias: +3000V
Design: Type I
Bias shutdown: ORTEC
Status:
|
Date |
Location |
Position |
Problems
|
REaction |
Comments |
Initials |
|
|
Hospital |
Annealing Station |
Neutron damage |
- |
- |
A.H. |
|
|
Hospital |
Testing |
Noisy channel |
- |
One side channel and one leaf (blue) are noisy. Needs further testing. Cold. |
A.H. |
|
|
SASSYER |
#1 |
Blue/side channel are still noisy |
- |
- |
A.H. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Detector E4
Serial Number: 16
Nominal Bias: +3500V
Design: Type I
Bias shutdown: ORTEC
Status:
|
Date |
Location |
Position |
Problems
|
REACTION |
Comments |
Initials |
|
|
Hospital |
- |
Bias shutdown shot? |
- |
Tom is working on it. Fixed in the evening. |
A.H. |
|
|
SPEEDY |
#2 |
- |
- |
- |
A.H. |
|
|
SASSYER |
#4 |
- |
- |
- |
A.H. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Detector E5
Serial Number:
Nominal Bias: V
Design: Type I
Bias shutdown: ORTEC
Status:
|
Date |
Location |
Position |
Problems
|
REACTION |
Comments |
Initials |
|
|
|
- |
Preamps dead |
- |
- |
A.H. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Detector E6
Serial Number: 19
Nominal Bias: +3000V
Design: Type I
Bias shutdown: ORTEC
Status:
|
Date |
Location |
Position |
Problems
|
REACTION |
Comments |
Initials |
|
|
SPEEDY |
#6 |
- |
- |
- |
A.H. |
|
|
SASSYER |
#6 |
- |
- |
- |
A.H. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Detector C7
Serial Number: 20
Nominal Bias: +3000V
Design: Type I
Bias shutdown: ORTEC
Status:
|
Date |
Location |
Position |
Problems
|
REACTION |
Comments |
Initials |
|
|
Hospital |
- |
Leaky cryostat |
- |
Needs to be fixed
by |
A.H. |
|
|
|
- |
- |
- |
Picked up today Z 04 11 851 |
A.H. |
|
|
Hospital |
- |
- |
- |
Back from |
A.H. |
|
|
Hospital |
- |
- |
- |
Cooling it down. |
A.H. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Detector E8
Serial Number: 23
Nominal Bias: +3000V
Design: Type I
Bias shutdown: ORTEC
Status:
|
Date |
Location |
Position |
Problems
|
REACTION |
Comments |
Initials |
|
|
SPEEDY |
#5 |
- |
- |
- |
A.H. |
|
|
SPEEDY |
#5 |
Warmed up |
- |
- |
A.H. |
|
|
Hospital |
- |
Neutron damage |
- |
Needs annealing |
A.H. |
|
|
Hospital |
Annealing Station |
Neutron Damage |
- |
- |
A.H. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Detector E9
Serial Number: 24
Nominal Bias: +3000V
Design: Type II
Bias shutdown: ORTEC
Status:
|
Date |
Location |
Position |
Problems
|
REACTION |
Comments |
Initials |
|
|
Hospital |
- |
Bias shutdown thresholds can’t be adjusted |
- |
Wait for word from |
A.H. |
|
|
|
- |
- |
- |
Picked up today Z 04 11 207 |
A.H. |
|
|
Hospital |
- |
fixed |
- |
Back from |
A.H. |
|
|
Hospital |
- |
- |
- |
Cooling it down. |
A.H. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Detector E10
Serial Number: 31
Nominal Bias: +4000V
Design: Type II
Bias shutdown: ORTEC
Status:
|
Date |
Location |
Position |
Problems
|
REACTION |
Comments |
Initials |
|
|
Hospital |
- |
Noise in side (right?) and black channels |
- |
Waits to be picked
up by |
A.H. |
|
|
Meriden/France |
- |
- |
- |
Picked up today Z 04 11 664 |
A.H. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Detector E11
Serial Number: 35
Nominal Bias: +3500V
Design: Type II
Bias shutdown: ORTEC
Values at delivery:
Measured efficiencies: Nominal:
Blue 28% 26.3%
Red 27.5% 26.4%
Green 28.2% 26.1%
Black 28% 26.3%
Measured Peak-to-Compton: Nominal:
Blue 51.3 54.6
Red 51.5 52.2
Green 52.4 53.1
Black 48.9 54.4
Resolution measured at 1.33 Mev:
Blue (FWHM/FWTM 1.90/3.56, DC offset 5mv ) Nominal: FWHM 1.86
Red (FWHM/FWTM 1.91/3.60, DC offset 2mv ) Nominal: FWHM 1.92
Green (FWHM/FWTM 1.93/3.54, DC offset 2mv ) Nominal: FWHM 1.89
Black (FWHM/FWTM 1.97/3.53, DC offset 2mv ) Nominal: FWHM 1.83
Left (FWHM/FWTM 4.10/6.60) Nominal:
FWHM 5.4
Right (FWHM/FWTM 3.53/5.62) Nominal:
FWHM 5.4
Middle (FWHM/FWTM 4.65/6.92) Nominal:
FWHM 7.9
Status:
|
Date |
Location |
Position |
Problems
|
REACTION |
Comments |
Initials |
|
|
- |
- |
- |
- |
Not yet delivered |
A.H. |
|
|
Hospital |
- |
- |
- |
Needs assembling, testing, cooling, changing of bias shutdown(?) |
A.H. |
|
|
Hospital |
- |
- |
- |
Getting cold |
A.H. |
|
|
SASSYER |
11 |
- |
- |
Tested and bias shutdown changed (it came with the wrong one). |
A.H. |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|