Autofill Page (internal)

 

Autofill person of the week starting November 8, 2004: John Ai

Autofill person of the week starting November 22, 2004: Gulhan Gurdal

Autofill person of the week starting November 29, 2004: Jing Qian

Autofill person of the week starting December 6, 2004: Andreas Heinz

Autofill person of the week starting December 13, 2004: Libby McCutchan

 

SPEEDY positions:

 

Starting with the upstream hemisphere and the top detector and count clockwise:

                                               #1

                                   #4         X         #2

                                               #3

with X being the beam into the page.

Correspondingly for the downstream hemisphere:

                                               #5

                                   #8         X         #6

                                               #7

 

SASSYER positions:

 

As numbered at the array.

 

 

Detector E1

Serial Number: 06

Nominal Bias: +3000 V

Design: Type I

Bias shutdown: ORTEC

Status:

Date

Location

Position

Problems

REaction

Comments

Initials

11/9/2004

SPEEDY

#3

-

-

-

A.H.

11/16/2004

SASSYER

#12

-

-

-

A.H.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Detector E2

Serial Number: 07

Nominal Bias: +3000V

Design: Type I

Bias shutdown: Ortec

Status:

Date

Location

Position

Problems

REACTION

Comments

Initials

11/9/2004

SPEEDY

#2

-

-

-

A.H.

11/10/2004

Hospital

-

-

-

Tom changes bias shutdown from Eurosys to Ortec.

Finished in the evening.

A.H.

11/16/2004

SASSYER

#4

-

-

-

A.H.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Detector E3

Serial Number: 10

Nominal Bias: +3000V

Design: Type I

Bias shutdown: ORTEC

Status:

Date

Location

Position

Problems

REaction

Comments

Initials

11/9/2004

Hospital

Annealing Station

Neutron damage

-

-

A.H.

11/15/2004

Hospital

Testing

Noisy channel

-

One side channel and one leaf (blue) are noisy. Needs further testing.

Cold.

A.H.

12/13/04

SASSYER

#1

Blue/side channel are still noisy

-

-

A.H.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Detector E4

Serial Number: 16

Nominal Bias: +3500V

Design: Type I

Bias shutdown: ORTEC

Status:

Date

Location

Position

Problems

REACTION

Comments

Initials

11/9/2004

Hospital

-

Bias shutdown shot?  

-

Tom is working on it.

Fixed in the evening.

A.H.

11/10/2004

SPEEDY

#2

-

-

-

A.H.

11/16/2004

SASSYER

#4

-

-

-

A.H.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Detector E5

Serial Number:

Nominal Bias: V

Design: Type I

Bias shutdown: ORTEC

Status:

Date

Location

Position

Problems

REACTION

Comments

Initials

11/9/2004

France

-

Preamps dead

-

-

A.H.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Detector E6

Serial Number:  19

Nominal Bias: +3000V

Design: Type I

Bias shutdown: ORTEC

Status:

Date

Location

Position

Problems

REACTION

Comments

Initials

11/9/2004

SPEEDY

#6

-

-

-

A.H.

11/16/2004

SASSYER

#6

-

-

-

A.H.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Detector C7

Serial Number: 20

Nominal Bias: +3000V

Design: Type I

Bias shutdown: ORTEC

Status:

Date

Location

Position

Problems

REACTION

Comments

Initials

11/9/2004

Hospital

-

Leaky cryostat

-

Needs to be fixed by Canberra

A.H.

11/17/2004

Meriden

-

-

-

Picked up today

Z 04 11 851

A.H.

12/08/2004

Hospital

-

-

-

Back from Meriden. Canberra claims that Cryostat is fixed.

A.H.

12/09/2004

Hospital

-

-

-

Cooling it down.

A.H.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Detector E8

Serial Number: 23

Nominal Bias: +3000V

Design: Type I

Bias shutdown: ORTEC

Status:

Date

Location

Position

Problems

REACTION

Comments

Initials

11/9/2004

SPEEDY

#5

-

-

-

A.H.

11/12/2004

SPEEDY

#5

Warmed up

-

-

A.H.

11/16/2004

Hospital

-

Neutron damage

-

Needs annealing

A.H.

12/6/2004

Hospital

Annealing Station

Neutron Damage

-

-

A.H.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Detector E9

Serial Number: 24

Nominal Bias: +3000V

Design: Type II

Bias shutdown: ORTEC

Status:

Date

Location

Position

Problems

REACTION

Comments

Initials

11/9/2004

Hospital

-

Bias shutdown thresholds can’t be adjusted

-

Wait for word from Canberra

A.H.

11/17/2004

Meriden

-

-

-

Picked up today

Z 04 11 207

A.H.

12/8/2004

Hospital

-

fixed

-

Back from Meriden

A.H.

12/9/2004

Hospital

-

-

-

Cooling it down.

A.H.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Detector E10

Serial Number: 31

Nominal Bias: +4000V

Design: Type II

Bias shutdown: ORTEC

Status:

Date

Location

Position

Problems

REACTION

Comments

Initials

11/9/2004

Hospital

-

Noise in side (right?) and black channels

-

Waits to be picked up by Canberra

A.H.

11/17/2004

Meriden/France

-

-

-

Picked up today

Z 04 11 664

A.H.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Detector E11

Serial Number:  35

Nominal Bias:  +3500V

Design:  Type II

Bias shutdown: ORTEC

Values at delivery:

Measured efficiencies:                           Nominal:

Blue      28%                                         26.3%  

Red      27.5%                                       26.4%

Green   28.2%                                       26.1%

Black    28%                                         26.3%

Measured Peak-to-Compton:                  Nominal:

Blue 51.3                                              54.6

Red 51.5                                               52.2

Green 52.4                                            53.1

Black 48.9                                             54.4

Resolution measured at 1.33 Mev:

Blue (FWHM/FWTM 1.90/3.56, DC offset 5mv )                           Nominal: FWHM 1.86 

Red (FWHM/FWTM 1.91/3.60, DC offset 2mv )                           Nominal: FWHM 1.92

Green (FWHM/FWTM 1.93/3.54, DC offset 2mv )                        Nominal: FWHM 1.89

Black (FWHM/FWTM 1.97/3.53, DC offset 2mv )                         Nominal: FWHM 1.83

Left (FWHM/FWTM 4.10/6.60)                                                   Nominal: FWHM 5.4

Right (FWHM/FWTM 3.53/5.62)                                                 Nominal: FWHM 5.4

Middle (FWHM/FWTM 4.65/6.92)                                               Nominal: FWHM 7.9

 

Status:

Date

Location

Position

Problems

REACTION

Comments

Initials

11/9/2004

-

-

-

-

Not yet delivered

A.H.

11/10/2004

Hospital

-

-

-

Needs assembling, testing, cooling,

changing of bias shutdown(?)

A.H.

11/17/2004

Hospital

-

-

-

Getting cold

A.H.

12/09/2004

SASSYER

11

-

-

Tested and bias shutdown changed (it came with the wrong one).

A.H.